Section 4: Configuration and Service Modes
4- 6
4.2.13 Configuration Menu
Entering the configuration menu steady state from the configuration mode
steady state, allows you to view the UIF software version, enable/disable the
measurement complete tone, or reset to factory power-on defaults. To return
to the configuration mode steady state from the configuration menu steady
state, while in the configuration mode steady state, press the LOWER
ALARM LIMIT button.
1.
From configuration mode steady state, press the UPPER and LOWER
ALARM LIMIT buttons simultaneously to access the configuration menu
steady state. The currently selected menu item number (1 to 3) is
displayed in the SYSTOLIC display.
2.
To view the software version of the UIF module, press the UP or DOWN
ARROW buttons until “1” is displayed in the SYSTOLIC display. Press
the UPPER ALARM LIMIT button. The PULSE RATE indicator will
light.
The digit in the left-hand display character of the SYSTOLIC display
represents the software version. The next two characters in the
SYSTOLIC display represent the major software revision. The three
characters in the DIASTOLIC display represent the minor software
revision. To return to the configuration menu steady state, press the
LOWER ALARM LIMIT button or allow the unit to time out (10
seconds).
3.
To enable or disable the measurement complete tone, press the UP or
DOWN ARROW buttons until “2” is displayed in the SYSTOLIC display.
Press the UPPER ALARM LIMIT button. The PULSE RATE indicator
will light. The current state is displayed in the DIASTOLIC display
(0 = disabled, 1 = enabled).
To change the current state, press the UP/DOWN ARROW buttons until
the desired state is displayed. To return to the configuration menu
steady state, press the LOWER ALARM LIMIT key or allow the unit to
time out (10 seconds).
4.
To reset institutional default parameters to factory default values, press
the UP ARROW button until “3” is displayed in the SYSTOLIC display.
Press the UPPER ALARM LIMIT button to reset the parameters. The
monitor automatically returns to the configuration menu steady state
when the values are reset.