![background image](http://html.mh-extra.com/html/nellcor/symphony-n-3100/symphony-n-3100_service-manual_718057016.webp)
Section 3: Performance Verification
3- 6
21. Press the ON/STANDBY button to turn the monitor off. Verify that the
monitor emits three consecutively lower-pitched beeps.
22. Press the ON/STANDBY button to turn the monitor on. The monitor
performs the Power-On Self-Test described in 3.3.1.1, steps 3a.
through 3g.
23. From the normal mode steady state, press and release the UPPER
ALARM LIMIT button. Verify that the monitor emits a single beep and
the SYSTOLIC display indicates an alarm limit of “240” for a few
seconds.
24. From the normal mode steady state, press and release the LOWER
ALARM LIMIT button. Verify that the monitor emits a single beep and
that the SYSTOLIC display indicates an alarm limit of “70” for a few
seconds.
25. From the normal mode steady state, press the UPPER ALARM LIMIT
button two times rapidly. Verify that the monitor emits two beeps and
that the DIASTOLIC display indicates an alarm limit of “210” for a few
seconds.
26. From the normal mode steady state, press the LOWER ALARM LIMIT
button two times rapidly. Verify that the monitor emits two beeps and
that the DIASTOLIC display indicates an alarm limit of “50” for a few
seconds.
27. From the normal mode steady state, press the UPPER ALARM LIMIT
button three times rapidly. Verify that the monitor emits three beeps
and that the MEAN/PULSE RATE display indicates an alarm limit of
“225” for a few seconds.
28. From the normal mode steady state, press the LOWER ALARM LIMIT
button three times rapidly. Verify that the monitor emits three beeps
and the MEAN/PULSE RATE display indicates an alarm limit of “55” for
a few seconds.
29. From the normal mode steady state, press the UPPER ALARM LIMIT
button four times rapidly. Verify that the monitor emits four beeps, the
PULSE RATE indicator lights and that the MEAN/PULSE RATE display
indicates an alarm limit of “170” for a few seconds.
30. From the normal mode steady state, press the LOWER ALARM LIMIT
button four times rapidly. Verify that the monitor emits four beeps, the
PULSE RATE indicator lights, and the MEAN/PULSE RATE display
indicates an alarm limit of “40” for a few seconds.
31. Press the ON/STANDBY button to place the N-3100 in the standby
mode. Verify that the monitor emits three consecutively higher-pitched
beeps.