Section 4: Configuration and Service Modes
4- 5
4.2.10 Alarm Volume
Use the following procedure to adjust the alarm volume default:
1.
From the configuration mode steady state, press and hold the ALARM
SILENCE button for approximately 3 seconds. A continuous tone at the
current volume setting is emitted.
2.
“VOL” is displayed in the DIASTOLIC display. A value from 1 to 10
(1 = lowest, 10 = highest) corresponding to the current volume setting is
displayed in the MEAN/PULSE RATE display.
3.
Continue to hold the ALARM SILENCE button and press the UP or
DOWN ARROW button to increase or decrease the volume.
4.
To exit this procedure and set the alarm volume default, release the
ALARM SILENCE button.
4.2.11 Alarm Silence Duration
Use the following procedure to adjust the alarm silence duration default:
1.
From the configuration mode steady state, press and hold the ALARM
SILENCE button. The current default setting, in seconds, for the silence
duration appears in the DIASTOLIC display. The SYSTOLIC display
indicates “CFG,” while the MEAN/PULSE RATE display indicates “SEC”
(seconds).
2.
Within 3 seconds and while holding the ALARM SILENCE button, press
the UP or DOWN ARROW button to adjust the silence duration. Silence
duration choices are 30 SEC, 60 SEC, 90 SEC, and 120 SEC.
3.
To exit this procedure and set the alarm silence duration default, release
the ALARM SILENCE button and return to the configuration mode
steady state.
4.2.12 Auto Measurement Cycle Interval
Use the following procedure to adjust the auto measurement cycle interval
default:
1.
From the configuration mode steady state, press and release the CYCLE
button. The current default setting, in minutes, for the auto
measurement cycle interval appears in the MEAN/PULSE RATE
display. The SYSTOLIC display indicates “CFG.” The CYCLE indicator
lights.
2.
Use the UP or DOWN arrows to adjust the cycle interval default. Cycle
interval choices are 1, 3, 5, 10, 15, 30, 60, 90 minutes and OFF.
3.
To exit this procedure and set the monitor to power-up in the manual
mode, select OFF as the cycle interval and allow the unit to time out and
return to the configuration mode steady state.
To exit this procedure and set the monitor to power-up in the automatic
mode, select a cycle interval other than OFF and allow the unit to time
out and return to the configuration mode steady state. The cycle interval
selected will be the default cycle interval.