Section 6: Disassembly Guide
6- 3
5.
Separate the bottom enclosure from the top half of the monitor, keeping
the right side panel attached to the bottom enclosure and the left side
panel attached to the top.
6.
Disconnect the blood pressure module ribbon connector from the UIF
board as illustrated in Figure 6-3.
Handle
Left panel
Bottom enclosure
Pump hose
ON/STBY button
Piezo
assembly
UIF board
Lower docking
connector ribbon
Blood Pressure module
ribbon connector
Right panel
Figure 6-3: Separating the Monitor and Disconnecting the Cables
6.3.1 Replacing the Piezo Speaker
To replace the Piezo speaker:
1.
Disconnect the Piezo assembly wires from the UIF board.
2.
Unsnap the speaker from the form-fitted side cover by gently prying it
with a small screwdriver.
3.
Install a new speaker.