Section 4: Configuration and Service Modes
4- 21
11. The first two digits of the DIASTOLIC display will be the fifth byte of the
version in hexadecimal format. The third digit will be “H.” The first two
digits of the MEAN/PULSE RATE display will be the sixth byte of the
version in hexadecimal format. The third digit will be “H.”
12. Press the UP ARROW button to display “37.5” in the SYSTOLIC display.
13. The first two digits of the DIASTOLIC display will be the seventh byte of
the version in hexadecimal format. The third digit will be “H.”
14. Press the LOWER ALARM LIMIT button to return to the service mode
steady state.
Caution: Menu items 70 and above are for factory purposes only.
Adjustment of menu items 70 and above by other than qualified
factory personnel may cause the N-3100 to malfunction.