Section 5: Troubleshooting
5- 10
Table 5-8: Stack Problems (Cont.)
7. While operating on
battery power, N-3000
operates with BATTERY
IN USE/BATTERY LOW
indicator lighting
steadily but N -3100 does
not operate.
Ensure that a good docking connection
exists between the N-3000 and N -3100.
8. Serial data or the alarm
active function from the
N-3100 is not available
at the serial data output
of the N-3000 or cannot
be displayed on the
N-3200.
1. Ensure that a good docking connection
exists between the N-3100 and the N -3000,
the N -3200, or both.
2. Check that an SPS power supply connected
to the N -3100 is functioning properly or
that AC power connected to the N-3200 is
correct.
9. The AUDIBLE ALARM
SILENCE button and
knob on an N -3000
stacked with the N-3100
have no effect on the
N-3100.
Earlier N -3000 versions do not have the
ability to communicate properly with the
N-3100. After ensuring that a good docking
connection exists, contact Nellcor Puritan
Bennett’s Technical Services Department or
your local Nellcor Puritant Bennett
representative for information on
retrofitting the N-3000.