SWG100
BIOGAS
Supplementary instruction:
External condensate trap
Version
01
Page 115
of
141
13.8.
Option: Extern condensate-trap
The external condensate trap is a pre-filter unit, which is for wet and dirty sampling points with
condensate higher than 14ml/min.
Mounting of the external condensate trap
The schematic above shows the tubing of the external condensate trap:
*Height of water column is dependent from the overpressure, which should be created.
1
mete
r=10
0h
Pa
*
Silicon hose: 6x2 mm
Area marked red: Must
be filled with water.
•
Silicon hose must be
laid as siphon.
•
The silicon hose must
be filled with water.
Hose holder
Condensate vessel:
Warning: Be sure, that the silicon
hose does not hang into the water
surface.
Sample gas inlet
Thread: 3x G1/8 inner
thread
Sample gas
outlet
PTFE-flexible tube: DN6/4 mm (max. 100m)
Safety valve (not scope of
supply).
Sample gas
inlet at the
analyzer
Fitting: Tube nipple:
For silicon hose 6x2
mm
Содержание SWG 100 biogas
Страница 1: ...MRU GmbH SWG 100 User Manual Page 1 of 141...
Страница 2: ...MRU GmbH SWG 100 User Manual Page 2 of 141...
Страница 13: ...MRU GmbH SWG 100 User Manual Page 13 of 141 Picture above Sample gas flow diagram...
Страница 81: ...MRU GmbH SWG 100 User Manual Page 81 of 141 Device info Path EXTRAS DEVICE INFO Structure...
Страница 86: ......
Страница 92: ......
Страница 133: ...MRU GmbH SWG 100 User Manual Page 133 of 141 13 15 Option NDIR bench for 0 3 000ppm CO2 0 CH4...
Страница 134: ...MRU GmbH SWG 100 User Manual Page 134 of 141 13 16 Option Dilution unit for H2S and CO electrochemical sensors...
Страница 140: ...MRU GmbH SWG 100 User Manual Page 140 of 141 14 Declaration of conformity...
Страница 141: ...MRU GmbH SWG 100 User Manual Page 141 of 141 V01 02 EN...