7.3.
Configuration of the alarm relays
On the main PCB there is o
contact. The following errors
will turn the relay from NC to NO.
1.
Main board is offline (internal RS485 bus communication failure)
2.
3.
Gas leakage inside analyzer cabinet (CH4 > 20% to 50% LEL)
4.
Condensate alarm (contacts resistance <
5.
Low fan rotation (speed rotation < 900U/min)
6.
Sample flow alarm (sample flow < 20 l/h)
7.
Gas cooler high alarm (temperature > +10°C)
8.
Gas cooler low alarm (temperature < +2°C)
9.
Cabinet high temperature (> +55°C)
10.
Cabinet low temperature (< +5°C)
Errors
1 to 5
alarm will force a measurement stop (all analog outputs are on hold or at 2mA,
depending on configuration).
Errors
5 to 10
will be displayed as warning message only; analog outputs of active sampling point
are live, all others are on hold.
ATTENTION
Analyzer system alarm relay is a potential-free contact, which max.
24VDC/VAC and a current of 1A (max.)
Connection of the alarm relay: Potential-free
Max. 24VDC/VAC, 1 A (max.)
Содержание SWG 100 biogas
Страница 1: ...MRU GmbH SWG 100 User Manual Page 1 of 141...
Страница 2: ...MRU GmbH SWG 100 User Manual Page 2 of 141...
Страница 13: ...MRU GmbH SWG 100 User Manual Page 13 of 141 Picture above Sample gas flow diagram...
Страница 81: ...MRU GmbH SWG 100 User Manual Page 81 of 141 Device info Path EXTRAS DEVICE INFO Structure...
Страница 86: ......
Страница 92: ......
Страница 133: ...MRU GmbH SWG 100 User Manual Page 133 of 141 13 15 Option NDIR bench for 0 3 000ppm CO2 0 CH4...
Страница 134: ...MRU GmbH SWG 100 User Manual Page 134 of 141 13 16 Option Dilution unit for H2S and CO electrochemical sensors...
Страница 140: ...MRU GmbH SWG 100 User Manual Page 140 of 141 14 Declaration of conformity...
Страница 141: ...MRU GmbH SWG 100 User Manual Page 141 of 141 V01 02 EN...