MRU GmbH
SWG 100 User Manual
Page
124
of
141
2.
Open the path Extras/General settings, at the analyzer. Push
F3=Modbus
.
At this menu the Modbus slave parameters can be configurated.
Example:
Baud rate: 19200
Slave address: 238
Stop bits: 1
Parity: even
Data bits: 8
Picture: The screenshot shows an example of a Modbus slave settings.
3.
In the next step it is necessary to build a connection between the analyzer
and the network. To do this open the browser at a pc and open the
webserver with the address:
•
http://192.xxx.xxx.xxx/maintenance/index.html
Important
With the test modus for the alarm outputs of the IO module, the alarm
relays at the IO modules and the relay at the PCB mainboard can be
activated.
Important
The 192.xxxx.xxx.xxx is the IP-address, which is configurated at the
module (see chapter xx).
Example:
The IP-address of the module is 192.168.100.154, then the correct address to invoke
the webserver is:
/maintenance/index.html.
4.
The webserver-address is protected with a password. The default password
is:
a.
Username: admin
b.
Password: admin.
5.
The following screen appears.
Содержание SWG 100 biogas
Страница 1: ...MRU GmbH SWG 100 User Manual Page 1 of 141...
Страница 2: ...MRU GmbH SWG 100 User Manual Page 2 of 141...
Страница 13: ...MRU GmbH SWG 100 User Manual Page 13 of 141 Picture above Sample gas flow diagram...
Страница 81: ...MRU GmbH SWG 100 User Manual Page 81 of 141 Device info Path EXTRAS DEVICE INFO Structure...
Страница 86: ......
Страница 92: ......
Страница 133: ...MRU GmbH SWG 100 User Manual Page 133 of 141 13 15 Option NDIR bench for 0 3 000ppm CO2 0 CH4...
Страница 134: ...MRU GmbH SWG 100 User Manual Page 134 of 141 13 16 Option Dilution unit for H2S and CO electrochemical sensors...
Страница 140: ...MRU GmbH SWG 100 User Manual Page 140 of 141 14 Declaration of conformity...
Страница 141: ...MRU GmbH SWG 100 User Manual Page 141 of 141 V01 02 EN...