Dilution-unit
The dilution-unit inside the analyser dilutes the sample gas flow with a fix ratio. The unit is
preadjusted. The operator must do a fine calibration of the device, to improve the accuracy of the
measurement. For the fine calibration it is indispensable to tube the vent and all sample gas inlets,
before making the actual calibration. The correct order for the fine calibration is shown in the flow
chart.
Correct installation of the SWG-100
Tubing of the vent
Tubing of all sample gas inlets
Installation of the calibration gas cylinder
Fine calibration of the dilution-unit