MRU GmbH
SWG 100 User Manual
Page
26
of
141
7.
Analyzer commissioning manual
After the first start of the analyzer it is necessary to make some settings at the analyzer. These
settings are:
•
Check the country and language.
•
Check the date and time of the instrument.
•
Configuration of the alarm relays.
•
Configuration of the Modbus.
•
Configuration of the external control via relay contacts (IO module).
•
Configuration of the analog outputs at the I/O module.
•
Configuration of the AUX inputs at the I/O module.
•
Configuration of the alarm outputs at the I/O module.
•
Configuration of the auto calibration.
7.1.
Check country and language
Important note:
In case the analyzer shows a language, you don't understand, you may swap the language to
English by pressing the menu key and selecting the function 'Set English language'.
Use the menu
E
XTRAS
G
ENERAL
S
ETTINGS
.
The analyser will automatically set some country-typical parameters like the language, the date
format, the temperature unit, the daylight-savings time function and the CSV-export settings.
7.2.
Check date and time of the instrument
The analyzer stores automatically measurement values including timestamps. Therefore the
instruments' system clock should be set correctly.
Use the menu
E
XTRAS
G
ENERAL
S
ETTINGS
D
ATE
&
T
IME
.
In case the date & time is not correct, press the key
F2
=modify, change date & time and then press
the key
F2
=store.
Note:
According to the selected country (see previous chapter) the analyzer automatically switches the
daylight saving time in spring and autumn. This function is active for most European countries.
Whenever the daylight-saving time is currently active, then you'll see a '*' in the time line of the
menu, thus 'Time *' instead of 'Time'.
Содержание SWG 100 biogas
Страница 1: ...MRU GmbH SWG 100 User Manual Page 1 of 141...
Страница 2: ...MRU GmbH SWG 100 User Manual Page 2 of 141...
Страница 13: ...MRU GmbH SWG 100 User Manual Page 13 of 141 Picture above Sample gas flow diagram...
Страница 81: ...MRU GmbH SWG 100 User Manual Page 81 of 141 Device info Path EXTRAS DEVICE INFO Structure...
Страница 86: ......
Страница 92: ......
Страница 133: ...MRU GmbH SWG 100 User Manual Page 133 of 141 13 15 Option NDIR bench for 0 3 000ppm CO2 0 CH4...
Страница 134: ...MRU GmbH SWG 100 User Manual Page 134 of 141 13 16 Option Dilution unit for H2S and CO electrochemical sensors...
Страница 140: ...MRU GmbH SWG 100 User Manual Page 140 of 141 14 Declaration of conformity...
Страница 141: ...MRU GmbH SWG 100 User Manual Page 141 of 141 V01 02 EN...