MRU GmbH
SWG 100 User Manual
Page
28
of
141
Plug connector definition for the system alarm relay
Slit screws
Stripping length:
7 mm
Tightening torque min.-max.:
0,5-0,6 Nm
Conductor cross sections, which can be used:
Type of electric line
Conductor cross section min.-max.
Solid
0,2-2,5 mm² (30-12 AWG)
Stranded
0,2-2,5 mm² (30-12 AWG)
Solid with ferrule (with/ or without plastic)
0,25-2,5 mm²
Information for cables, which go through the cable gland M16:
It is recommended to use only electric lines with ferrules.
Following analyzer errors will produce a system alarm (open contact of System Alarm relay)