Example: 4-20 mA signal outputs to control room
Starting position:
A plan wants to read the measurement values from the CO2 channel to their DCS
1
.
The SWG100 has only one IO module and three sample points. All outputs are free.
The CO2 values from the second sample point should be logged in the control
room.
Follow steps must be done to do this:
1.
Wiring the analog-output 1 with the DCS: Wiring the analog-output
(AO1+/AO1-) with the DCS. For this, use a wiring with a conductor cross section
between
2.
In the next step output 1 must be configured at the SWG100. To do this, open
the menu:
EXTRAS/ CONFIGURATE ANALOG OUTPUTS
. The following
menu overview will be appeared (see screenshot below). To configure the IO1,
select the first record (channel I/O 1/1 SP1 / CH4) in the list and open it.
1
DCS=Distributed Control System
DCS
Содержание SWG 100 biogas
Страница 1: ...MRU GmbH SWG 100 User Manual Page 1 of 141...
Страница 2: ...MRU GmbH SWG 100 User Manual Page 2 of 141...
Страница 13: ...MRU GmbH SWG 100 User Manual Page 13 of 141 Picture above Sample gas flow diagram...
Страница 81: ...MRU GmbH SWG 100 User Manual Page 81 of 141 Device info Path EXTRAS DEVICE INFO Structure...
Страница 86: ......
Страница 92: ......
Страница 133: ...MRU GmbH SWG 100 User Manual Page 133 of 141 13 15 Option NDIR bench for 0 3 000ppm CO2 0 CH4...
Страница 134: ...MRU GmbH SWG 100 User Manual Page 134 of 141 13 16 Option Dilution unit for H2S and CO electrochemical sensors...
Страница 140: ...MRU GmbH SWG 100 User Manual Page 140 of 141 14 Declaration of conformity...
Страница 141: ...MRU GmbH SWG 100 User Manual Page 141 of 141 V01 02 EN...