MRU GmbH
SWG 100 User Manual
Page
59
of
141
NOTE
All gas concentrations are in percent! The factory from percent to ppm is:
1%= 10.000 ppm.
5.
The user has the following options:
- leave the menu: The set gas concentrations are stored and the auto calibration will start after
the interval is reached.
- Start the auto- calibration immediately: Press
F2
(start now).
6.
During the auto calibration the stored gas concentrations are compared with the measure
gas concentrations. If the gas concentrations are not differing too much, the measured
7.
The new gas factors will be stored, after successful calibration.