MRU GmbH
SWG 100 User Manual
Page
101
of
141
•
Defined registers to be read by the master.
modbus
address
profibus
address
data
type
register content
Status & Device info
0
0
U32
Analyser Status
(more details read below)
2
4
U32
System Alarm
(more details read below)
4
8
U32
Serial number
6
12
U32
Analyser type
(11060 = SWG100biogas)
8
16
U32
Firmware version
(e.g. 12345 = V1.23.45)
10
20
U32
Elapsed seconds since Power-On
12
24
U32
Counter Modbus Frame Error
14
28
FL
CH4 amb. [%]
16
32
FL
CH4 amb. [% LEL]
18
36
FL
T-sensor [°C/°F]
(unit depends on user settings)
20
40
FL
Sample Flow [l/h]
22
44
FL
T-gascooler [°C/°F]
(unit depends on user settings)
24
48
FL
Case fan rotations [rpm]
26
52
FL
Gas pump rotations [rpm]
28
56
FL
P-absolute [hPa]
(=[mbar]
)
30
60
FL
P-absolute [inchHG]
32
U32
not (yet) defined
(reads zero)
34
U32
not (yet) defined
(reads zero)
36
U32
not (yet) defined
(reads zero)
38
U32
not (yet) defined
(reads zero)
Status & current measurement values (live values!)
40
U32
Analyser Status
(more details read below)
42
U32
System Alarm
(more details read below)
44
FL
O2 [%]
46
FL
CO2 [%]
(or CO2 [ppm], if CO2 [%] isn't available)
48
FL
CH4 [%]
(or CH4 [ppm], if CH4 [%] isn't available)
50
FL
H2S [ppm]
(optional)
52
FL
H2 [ppm]
(optional)
54
FL
Net calorific value [MJ/kg]
56
FL
Gross calorific value [MJ/kg]
modbus
address
profibus
address
data
type
register content
58
FL
Net calorific value [MJ/m³]
60
FL
Gross calorific value [MJ/m³]
62
FL
CO [ppm]
(optional)
64
FL
CH4 [ppm]
(optional, when CH4 [%] also available)
66
FL
CO2 [ppm]
(optional, when CO2 [%] also available)
68
U32
not (yet) defined
(reads zero)
Status & measurement values of sample point 1
70
U32
Analyser Status
(more details read below)
72
U32
System Alarm
(more details read below)
74
64
FL
O2 [%]
76
68
FL
CO2 [%]
(or CO2 [ppm], if CO2 [%] isn't available)
78
72
FL
CH4 [%]
(or CH4 [ppm], if CH4 [%] isn't available)
80
76
FL
H2S [ppm]
(optional)
Содержание SWG 100 biogas
Страница 1: ...MRU GmbH SWG 100 User Manual Page 1 of 141...
Страница 2: ...MRU GmbH SWG 100 User Manual Page 2 of 141...
Страница 13: ...MRU GmbH SWG 100 User Manual Page 13 of 141 Picture above Sample gas flow diagram...
Страница 81: ...MRU GmbH SWG 100 User Manual Page 81 of 141 Device info Path EXTRAS DEVICE INFO Structure...
Страница 86: ......
Страница 92: ......
Страница 133: ...MRU GmbH SWG 100 User Manual Page 133 of 141 13 15 Option NDIR bench for 0 3 000ppm CO2 0 CH4...
Страница 134: ...MRU GmbH SWG 100 User Manual Page 134 of 141 13 16 Option Dilution unit for H2S and CO electrochemical sensors...
Страница 140: ...MRU GmbH SWG 100 User Manual Page 140 of 141 14 Declaration of conformity...
Страница 141: ...MRU GmbH SWG 100 User Manual Page 141 of 141 V01 02 EN...