Fine calibration of the dilution-unit
If the steps above are done, the device should look like in the sketch below:
Follow steps must be done, for the fine calibration of the dilution-unit. The steps
below show the entire adjustment on the example of a H2S-sensors.
Max. x10 gas inlets