MRU GmbH
SWG 100 User Manual
Page
45
of
141
8.5.
Analog inputs (4-20 mA)
Hardware-side:
The analog inputs are on the top of the IO module. They are marked with a blue
label. Through the help of the analog inputs, the IO module can read all common 4-
20 mA transducer in the analyzer directly. The IO module has a separate 12 V power
source, for the supply of the transducer. At the sketch below, the inputs and the
power supply, for the transducer are marked.
Information:
Internal Load:
50 Ohm
Power supply for transducer:
12VDC / 200 mA
The sketches below show how the common transducer can connect to the IO
module.
V12+
AI1
+
AI2
+
AI1
-
AI2
-
AI3
+
AI4
+
AI3
-
AI4
-
Power supply for transducer
Analog inputs
GND
Содержание SWG 100 biogas
Страница 1: ...MRU GmbH SWG 100 User Manual Page 1 of 141...
Страница 2: ...MRU GmbH SWG 100 User Manual Page 2 of 141...
Страница 13: ...MRU GmbH SWG 100 User Manual Page 13 of 141 Picture above Sample gas flow diagram...
Страница 81: ...MRU GmbH SWG 100 User Manual Page 81 of 141 Device info Path EXTRAS DEVICE INFO Structure...
Страница 86: ......
Страница 92: ......
Страница 133: ...MRU GmbH SWG 100 User Manual Page 133 of 141 13 15 Option NDIR bench for 0 3 000ppm CO2 0 CH4...
Страница 134: ...MRU GmbH SWG 100 User Manual Page 134 of 141 13 16 Option Dilution unit for H2S and CO electrochemical sensors...
Страница 140: ...MRU GmbH SWG 100 User Manual Page 140 of 141 14 Declaration of conformity...
Страница 141: ...MRU GmbH SWG 100 User Manual Page 141 of 141 V01 02 EN...