MRU GmbH
SWG 100 User Manual
Page
79
of
141
Default configuration
Use the menu
M
EASUREMENT
->
E
XTRAS
>
D
EFAULT
S
ETTINGS
for the default configuration:
DEFAULT CONFIGURATION
Parameter
Analog output 4-
20mA
Alarm output relay contact
4mA
20mA
open
CH4 [ % ]
0
100
less than
50
O2 [ % ]
0
25
more than
1
H2S [ ppm ]
0
1000
more than
300
CO2 [ % ]
0
100
more than
50
H2 [ ppm ]
0
500
more than
500
0
500
more than
500
CH4 ambient [ % ]
0
5
more than
1
CH4 ambient [ %LEL ]
0
100
more than
20
Temperature sensor [ °C ]
0
50
more than
50
Fan rotation [ rpm ]
0
2000
less than
1200
Pump rotation [ rpm ]
0
5000
less than
1500
Sample flow [ l/hr ]
0
60
less than
30
Temperature cooler [°C ]
0
20
more than
10
Net calorific value [MJ/kg ]
0
40
less than
30
Gross calorific value [ MJ/kg ]
0
40
less than
30
Net calorific value [MJ/m3]
0
40
less than
30
Gross calorific value [ MJ/m3]
0
40
less than
30
Содержание SWG 100 biogas
Страница 1: ...MRU GmbH SWG 100 User Manual Page 1 of 141...
Страница 2: ...MRU GmbH SWG 100 User Manual Page 2 of 141...
Страница 13: ...MRU GmbH SWG 100 User Manual Page 13 of 141 Picture above Sample gas flow diagram...
Страница 81: ...MRU GmbH SWG 100 User Manual Page 81 of 141 Device info Path EXTRAS DEVICE INFO Structure...
Страница 86: ......
Страница 92: ......
Страница 133: ...MRU GmbH SWG 100 User Manual Page 133 of 141 13 15 Option NDIR bench for 0 3 000ppm CO2 0 CH4...
Страница 134: ...MRU GmbH SWG 100 User Manual Page 134 of 141 13 16 Option Dilution unit for H2S and CO electrochemical sensors...
Страница 140: ...MRU GmbH SWG 100 User Manual Page 140 of 141 14 Declaration of conformity...
Страница 141: ...MRU GmbH SWG 100 User Manual Page 141 of 141 V01 02 EN...