ZS3 Service Manual
Page 146 of 295
Figure 14.13-ZS3
Notes: Screw tightening torque: M4: 6-7 kgf.cm
For upgrade gel warmer
Tools: Screw Driver,
Auxiliary material: Cable Tie, PVC electric insulation tape
1. Drop telescoping cover to find preinstalled printer power and USB cables. Cut cable-
tie to release their free ends and route them forward.
Figure 14.14-ZS3
2. Paste 70-80mm length of PVC electric insulation tape on the edge of the Stem
weldment
Figure 14.15-ZS3
3. Use a screwdriver to install the 1 Pan head screw M4X12 that attach the Gel warmer
cable to the Stem weldment
Wind a circle and Tie
the tail of the Gel
Warmer cable and
printer USB cable (If
printer with no
options)
Cut cable tie
Paste PVC tape
Содержание Zonare ZS3
Страница 1: ...ZS3 Diagnostic Ultrasound System Service Manual ...
Страница 34: ...ZS3 Service Manual Page 34 of 295 4 System Overview Major System Assemblies Figure 4 1 ZS3 System ...
Страница 120: ...ZS3 Service Manual Page 120 of 295 FTP Setup Enters Setup Network FTP ...
Страница 124: ...ZS3 Service Manual Page 124 of 295 ...
Страница 131: ...ZS3 Service Manual Page 131 of 295 13 System Diagrams ...
Страница 132: ...ZS3 Service Manual Page 132 of 295 Power Block Diagram Figure 13 1 ZS3 Power Block Diagram ...
Страница 133: ...ZS3 Service Manual Page 133 of 295 Cabling Diagram Figure 13 2 ZS3 Cabling Diagram ...
Страница 134: ...ZS3 Service Manual Page 134 of 295 System Block Diagram Figure 13 3 ZS3 Overall System Block Diagram ...
Страница 138: ...ZS3 Service Manual Page 138 of 295 Figure 14 6 ZS3 ...
Страница 157: ...ZS3 Service Manual Page 157 of 295 8 The system performs the second automatic reboot and enters the following page ...
Страница 185: ...ZS3 Service Manual Page 185 of 295 17 Preventative Maintenance Forms ...
Страница 217: ...ZS3 Service Manual Page 217 of 295 Figure 18 8 ZS3 ...
Страница 252: ...ZS3 Service Manual Page 252 of 295 Figure 19 43 ZS3 Power Cable USB Cable ...
Страница 295: ...P N 046 014026 00 2 0 ...