ZS3 Service Manual
Page 21 of 295
Total Recall Data (TRI) should not be relied on for primary storage of
diagnostic data. Use of TRI across software versions is not guaranteed.
The system is not indicated for differentiation between malignant and benign
breast lesions.
Precautions
– USB Memory Stick
USB Memory Sticks that are purchased from outside sources (besides
Mindray/Zonare) may not be compatible for use with the system.
Precautions
– TEE Transducer
TEE: Perform an electrical leakage test prior to each use of the transducer.
See the TEE Maintenance Guide or the TEE Leakage Test Quick Reference
Guide for details.
TEE: Examine the transducer prior to each use. See the TEE Maintenance
Guide or the TEE Quick Reference Guide for details.
TEE: Clean and disinfect after each use.
TEE: Always use a bite guard.
TEE: Always use single-use probe sheath.
TEE: Ensure articulation locks are disengaged and straighten mechanism
during device insertion or extraction from patient.
TEE: Remove the transducer from the patient prior to defibrillation.
TEE: Do not store the disinfected transducer in the carrying case. Only use
the case for transportation.
TEE transducer is a delicate medical instrument, handle with care.
The multiplane TEE probe is a precision instrument, which must be handled
with care. It may be damaged when dropped or abused. In particular, do not
allow the ultrasonic window in the tip to come into contact with a sharp object.
Do not touch this window unnecessarily. Never exert force onto the acoustic
window.
The transducer connector is not watertight, and should always be kept dry.
The control handle, although spray-watertight, should not be immersed.
This equipment contains no operator serviceable components. To prevent
electric shock, do not remove any covers or panels.
Never manually deflect the distal tip of the probe; use only the deflection
control wheels.
Do not kink, tightly coil, or apply excessive force on the probe cable or shaft.
Insulation failure may result.
Under normal conditions at full acoustic power the temperature of the tip does
not exceed 42.7°C. Be sure to check at least monthly that the temperature
increase of the tip is within limits:
Connect the probe to the Ultrasound system.
Adjust the acoustic power to the highest value possible.
Select Color Doppler mode.
Wait for 2 minutes.
Feel at the distal end of the probe if there is a temperature increase that
could be harmful for the patient.
Содержание Zonare ZS3
Страница 1: ...ZS3 Diagnostic Ultrasound System Service Manual ...
Страница 34: ...ZS3 Service Manual Page 34 of 295 4 System Overview Major System Assemblies Figure 4 1 ZS3 System ...
Страница 120: ...ZS3 Service Manual Page 120 of 295 FTP Setup Enters Setup Network FTP ...
Страница 124: ...ZS3 Service Manual Page 124 of 295 ...
Страница 131: ...ZS3 Service Manual Page 131 of 295 13 System Diagrams ...
Страница 132: ...ZS3 Service Manual Page 132 of 295 Power Block Diagram Figure 13 1 ZS3 Power Block Diagram ...
Страница 133: ...ZS3 Service Manual Page 133 of 295 Cabling Diagram Figure 13 2 ZS3 Cabling Diagram ...
Страница 134: ...ZS3 Service Manual Page 134 of 295 System Block Diagram Figure 13 3 ZS3 Overall System Block Diagram ...
Страница 138: ...ZS3 Service Manual Page 138 of 295 Figure 14 6 ZS3 ...
Страница 157: ...ZS3 Service Manual Page 157 of 295 8 The system performs the second automatic reboot and enters the following page ...
Страница 185: ...ZS3 Service Manual Page 185 of 295 17 Preventative Maintenance Forms ...
Страница 217: ...ZS3 Service Manual Page 217 of 295 Figure 18 8 ZS3 ...
Страница 252: ...ZS3 Service Manual Page 252 of 295 Figure 19 43 ZS3 Power Cable USB Cable ...
Страница 295: ...P N 046 014026 00 2 0 ...