ZS3 Service Manual
Page 150 of 295
Figure 14.24-ZS3
For upgrade gel warmer holder
Auxiliary material: Cable Tie, and PVC electric insulation tape
1. Tie Gel warmer cable together with Monitor power cable on front of the Stem
weldment. Replace telescoping cover. If printer is optional, tie printer power cable
and printer USB cable. Replace telescoping cover
Figure 14.25-ZS3
2. Thread Gel warmer cable through UI hole and connect it to Gel warmer. Use cable
tie to fix the Gel warmer cable to cable hanger. Notes: cable tie needs to thread
through the mounting hole in connector of Gel warmer cable.
Figure 14.26-ZS3
Cut cable tie
Cut cable tie
Gel warmer cable
Cable tie needs to thread
through the mounting hole in
connector of Gel warmer
cable
Cable tie
Содержание Zonare ZS3
Страница 1: ...ZS3 Diagnostic Ultrasound System Service Manual ...
Страница 34: ...ZS3 Service Manual Page 34 of 295 4 System Overview Major System Assemblies Figure 4 1 ZS3 System ...
Страница 120: ...ZS3 Service Manual Page 120 of 295 FTP Setup Enters Setup Network FTP ...
Страница 124: ...ZS3 Service Manual Page 124 of 295 ...
Страница 131: ...ZS3 Service Manual Page 131 of 295 13 System Diagrams ...
Страница 132: ...ZS3 Service Manual Page 132 of 295 Power Block Diagram Figure 13 1 ZS3 Power Block Diagram ...
Страница 133: ...ZS3 Service Manual Page 133 of 295 Cabling Diagram Figure 13 2 ZS3 Cabling Diagram ...
Страница 134: ...ZS3 Service Manual Page 134 of 295 System Block Diagram Figure 13 3 ZS3 Overall System Block Diagram ...
Страница 138: ...ZS3 Service Manual Page 138 of 295 Figure 14 6 ZS3 ...
Страница 157: ...ZS3 Service Manual Page 157 of 295 8 The system performs the second automatic reboot and enters the following page ...
Страница 185: ...ZS3 Service Manual Page 185 of 295 17 Preventative Maintenance Forms ...
Страница 217: ...ZS3 Service Manual Page 217 of 295 Figure 18 8 ZS3 ...
Страница 252: ...ZS3 Service Manual Page 252 of 295 Figure 19 43 ZS3 Power Cable USB Cable ...
Страница 295: ...P N 046 014026 00 2 0 ...