C - 10
BeneVision N1 Patient Monitor Operator’s Manual
Art-S Extreme
Alarm switch
Off
High limit
Adult: 175 mmHg
Pediatric: 130 mmHg
Neonate: 95 mmHg
Low limit
Adult: 75 mmHg
Pediatric: 60 mmHg
Neonate: 50 mmHg
Priority
High
Alarm Outputs
Off
Art-D Extreme
Alarm switch
Off
High limit
Adult: 105 mmHg
Pediatric: 80 mmHg
Neonate: 65 mmHg
Low limit
Adult: 35mmHg
Pediatric: 30 mmHg
Neonate: 15 mmHg
Priority
High
Alarm Outputs
Off
Art-M Extreme
Alarm switch
Off
High limit
Adult: 125 mmHg
Pediatric: 100 mmHg
Neonate: 75 mmHg
Low limit
Adult: 55 mmHg
Pediatric: 40 mmHg
Neonate: 30 mmHg
Priority
High
Alarm Outputs
Off
CPP
Alarm switch
On
High limit
Adult: 130 mmHg
Pediatric: 100 mmHg
Neonate: 90 mmHg
Low limit
Adult: 50 mmHg
Pediatric: 40 mmHg
Neonate: 30 mmHg
Priority
Med
Alarm Outputs
Off
Measure (for P1, P2)
All
Measure (for P3, P4)
Mean only
Sensitivity
Med
Speed
25 mm/sec
Item
Default Setting