BeneVision N1 Patient Monitor Operator’s Manual
C - 7
NIBP-D Alarm
switch
On
High limit
Adult: 90 mmHg
Pediatric: 70 mmHg
Neonate: 60 mmHg
Low limit
Adult: 50 mmHg
Pediatric: 40 mmHg
Neonate: 20 mmHg
Priority
Med
Alarm Outputs
Off
NIBP-M
Alarm switch
On
High limit
Adult: 110 mmHg
Pediatric: 90 mmHg
Neonate: 70 mmHg
Low limit
Adult: 60 mmHg
Pediatric: 50 mmHg
Neonate: 25 mmHg
Priority
Med
Alarm Outputs
Off
NIBP-S Extreme
Alarm switch
Off
High limit
Adult: 175 mmHg
Pediatric: 130 mmHg
Neonate: 95 mmHg
Low limit
Adult: 75 mmHg
Pediatric: 60 mmHg
Neonate: 35 mmHg
Priority
High
Alarm Outputs
Off
NIBP-D Extreme
Alarm switch
Off
High limit
Adult: 105 mmHg
Pediatric: 80 mmHg
Neonate: 65 mmHg
Low limit
Adult: 35 mmHg
Pediatric: 30 mmHg
Neonate: 15 mmHg
Priority
High
Alarm Outputs
Off
NIBP-M Extreme
Alarm switch
Off
High limit
Adult: 125 mmHg
Pediatric: 100 mmHg
Neonate: 75 mmHg
Low limit
Adult: 45 mmHg
Pediatric: 40 mmHg
Neonate: 20 mmHg
Priority
High
Alarm Outputs
Off
Item
Default Setting
Содержание BeneVision N1