A - 16
BeneVision N1 Patient Monitor Operator’s Manual
A.14.5
Temp Specifications
A.14.6
NIBP Specifications
Standard
Meet the standard of ISO 80601-2-56 2009
Technique
Thermal resistance
Operating mode
Direct mode
Measurement range
0 to 50 °C (32 to 122 °F)
Resolution
0.1°C
Accuracy
±0.1 °C or ±0.2 °F (excluding probe error)
Refreshing rate
≤1 s
Minimum time for accurate
measurement
Body surface: <100 s
Body cavity: <80 s
Alarm limit
Range
Step
T1/T2 High
(low limit +1.0) to 50.0 °C
(low limit +2.0) to 122.0 °F
0.1 °C
0.1 °F
T1/T2 Low
0.1 to (high limit - 1.0) °C
32.2 to (high limit - 2.0) °F
∆
T High
0.1 to 50.0 °C
0.2 to 90.0 °F
Standard
Meet standard of IEC 80601-2-30 2013
Technique
Oscillometry
Mode of operation
Manual, Auto, STAT, Sequence
Auto mode repetition intervals
1, 2, 2.5, 3, 5, 10, 15, 20, 30, 60, 90, 120, 180, 240 or 480 min
STAT mode cycle time
5 min
Max measurement time
Adult, pediatric: 180 s
Neonate: 90 s
Heart rate range
30 to 300 bpm
Measurement ranges
(mmHg)
Adult
Pediatric
Neonate
Systolic:
25 to 290
25 to 240
25 to 140
Diastolic:
10 to 250
10 to 200
10 to 115
Mean:
15 to 260
15 to 215
15 to 125
Accuracy
Max mean error: ±5 mmHg
Max standard deviation: 8 mmHg
Resolution
1mmHg
Initial cuff inflation pressure range
(mmHg)
Adult: 80 to 280
Pediatric: 80 to 210
Neonate: 60 to 140
Default initial cuff inflation pressure
(mmHg)
Adult: 160
Pediatric: 140
Neonate: 90
Software overpressure protection
Adult: 297±3 mmHg
Pediatric: 297±3 mmHg
Neonate: 147±3 mmHg