A - 20
BeneVision N1 Patient Monitor Operator’s Manual
Response time
For CO
2
measurements (using external CO
2
module without O
2
sensor):
Measured with a DRYLINE II neonatal watertrap and a 2.5-meter neonatal sampling line:
≤5.0 s @ 70 ml/min
Measured with a DRYLINE II neonatal watertrap and a 2.5-meter neonatal sampling line:
≤4.5 s @ 90 ml/min
Measured with a DRYLINE II adult watertrap and a 2.5-meter adult sampling line:
≤5.0 s @ 120 ml/min
For CO
2
measurements (using external CO
2
module with O
2
sensor):
Measured with a DRYLINE II neonatal watertrap and a 2.5-meter neonatal sampling line:
≤4.5 s @ 90 ml/min
Measured with a DRYLINE II adult watertrap and a 2.5-meter adult sampling line:
≤5.0 s @ 120 ml/min
For CO
2
measurements (using internal CO
2
module):
Measured with a standard Oridion sampling line:
≤5.0 s@50 ml/min
Measured with a prolonged Oridion sampling line:
<6.5 s@50 ml/min
For O
2
measurements:
Measured with a DRYLINE II neoonatal watertrap and a 2.5-meter neonatal sampling
line:
≤4.5 s @ 90 ml/min
Measured with a DRYLINE II adult watertrap and a 2.5-meter adult sampling line:
≤5.0 s @ 120 ml/min
Rise time
For CO
2
measurements (using external CO
2
module without O
2
sensor):
Measured with a DRYLINE II neonatal watertrap and a 2.5-meter neonatal sampling line:
≤250 ms@70 ml/min.
Measured with a DRYLINE II neonatal watertrap and a 2.5-meter neonatal sampling line:
≤250 ms@90 ml/min.
Measured with a DRYLINE II adult watertrap and a 2.5-meter adult sampling line:
≤300 ms@120 ml/min
For CO
2
measurements (using external CO
2
module with O
2
sensor):
Measured with a DRYLINE II neonatal watertrap and a 2.5-meter neonatal sampling line:
≤250 ms@90 ml/min.
Measured with a DRYLINE II adult watertrap and a 2.5-meter adult sampling line:
≤300 ms@120 ml/min
For CO
2
measurements (using internal CO
2
module):
Measured with a standard Oridion sampling line:
≤250 ms@50 ml/min
Measured with a prolonged Oridion sampling line:
≤280 ms@50 ml/min
For O
2
measurements:
Measured with a DRYLINE II neonatal watertrap and a 2.5-meter neonatal sampling line:
≤800 ms@90 ml/min.
Measured with a DRYLINE II adult watertrap and a 2.5-meter adult sampling line:
≤750 ms@120 ml/min
awRR measurement range
0 to 150 rpm
awRR measurement precision
<60 rpm: ±1 rpm
60 to 150 rpm: ±2 rpm
awRR resolution
1 rpm
Data sample rate
50 Hz