BeneVision N1 Patient Monitor Operator’s Manual
A - 17
*Measurement accuracy verification: In adult and pediatric modes, the blood pressure measurements measured
with this device are in compliance with the Standard for Non-invasive sphygmomanometers (ISO 81060-2)in
terms of mean error and standard deviation by comparing with intra-arterial or auscultatory measurements
(depending on the configuration) in a typical patient population. For auscultatory reference, the 5th Korotkoff
sound was used to determine the diastolic pressure.
In neonatal mode, the blood pressure measurements measured with this device are in compliance with the
American National Standard for Non-invasive sphygmomanometers (ISO 81060-2) in terms of mean error and
standard deviation by comparing with intra-arterial measurements (depending on the configuration) in a typical
patient population.
A.14.7
IBP Specifications
Static pressure measurement range
0 mmHg to 300 mmHg
Static pressure measurement accuracy
±3 mmHg
Alarm limit
Range (mmHg)
Step (mmHg)
NIBP-S High
Adult: (low limit + 5) to 265
Pediatric: (low limit + 5) to 195
Neonate: (low limit + 5) to 130
NIBP ≤ 50: 1
NIBP > 50: 5
NIBP-S Low
Adult: 41 to (high limit - 5)
Pediatric: 41 to (high limit - 5)
Neonate: 40 to (high limit - 5)
NIBP-M High
Adult: (low limit + 5) to 225
Pediatric: (low limit + 5) to 160
Neonate: (low limit + 5) to 105
NIBP-M Low
21 to (high limit - 5)
NIBP-D High
Adult: (low limit + 5) to 205
Pediatric: (low limit + 5) to 145
Neonate: (low limit + 5) to 95
NIBP-D Low
11 to (high limit - 5)
NIBP-S Extreme High
Adult: (NIBP-S high limit + 5) to 270
Pediatric: (NIBP-S high limit +5) to 200
Neonate: (NIBP-S high limit +5) to 135
NIBP ≤ 50: 1
NIBP > 50: 5
NIBP-S Extreme Low
Adult: 40 to (NIBP-S low limit - 5)
Pediatric: 40 to (NIBP-S low limit - 5)
Neonate: 30 to (NIBP-S low limit - 5)
NIBP-M Extreme High
Adult: (NIBP-M high limit + 5) to 230
Pediatric: (NIBP-M high limit + 5) to 165
Neonate: (NIBP-M high limit + 5) to 110
NIBP-M Extreme Low
20 to (NIBP-M low limit - 5)
NIBP-D Extreme High
Adult: (NIBP-D high limit + 5) to 210
Pediatric: (NIBP-D high limit + 5) to 150
Neonate: (NIBP-D high limit + 5) to 100
NIBP-D Extreme Low
10 to (NIBP-D low limit - 5)
Standard
Meet the standard of IEC 60601-2-34 2011.
Technique
Direct invasive measurement
IBP
Measurement range
-50 to 300 mmHg
Resolution
1 mmHg