C - 12
BeneVision N1 Patient Monitor Operator’s Manual
C.7.2
Sidestream CO
2
Default Settings
C.7.3
Microstream CO
2
Default Settings
C.7.4
Mainstream CO
2
Default Settings
Speed
6.25 mm/s
Scale
50 mmHg
Waveform Type
Draw
Item
Default Setting
Item
Default Setting
EtO
2
Alarm
switch
On
High limit
88%
Low limit
18%
Priority
Med
Alarm Outputs
Off
FiO
2
Alarm
switch
On
High limit
Adult and pediatric: 100%
Neonate: 90%
Low limit
18%
Priority
Med
Alarm Outputs
Off
BTPS Compensation
Off
O
2
Compensation
OR: 100%
Other departments: 21%
N
2
O Compenation
0%
AG Compenation
0%
Auto Standby
60 min
Operating Mode
Measure
Item
Default Setting
BTPS Compen
Off
Maximum Hold
20 sec
Auto Standby
Off
Operating Mode
Measure
Item
Default Setting
Maximum Hold
1
0 sec
O
2
Compensation
Off