BeneVision N1 Patient Monitor Operator’s Manual
2.3.2
Left View
NOTE
•
Dry the Multi-pin connector of the N1 before connecting the N1 to the Modular Rack or Dock in case
of water spray.
(1)
Alarm lamp:
When a physiological alarm or technical alarm occurs, this lamp lights and flashes corresponding with the alarm
priority:
•
High priority alarms: the lamp quickly flashes red.
•
Medium priority alarms: the lamp slowly flashes yellow.
•
Low priority alarms: the lamp lights in cyan without flashing.
(2)
Ambient light sensor
When screen brightness is set to auto, the system automatically adjusts screen brightens according to the
strength of ambient light.
(3)
Display
(4) Battery
LED
(5) External
power
LED
•
On: when external power supply is connected.
•
Off: when external power supply is not connected.
(1)
Communication interface: used for communication between the
N1 and host monitor (BeneVision N series monitor).
(2)
Infrared filter: used for communication between the N1 and
BeneView T series monitor; used for communication between the
N1 and N series monitor if the communication interface does not
work.
(3)
Contact: used for receiving power supply from the host monitor
(BeneView T series monitor or BeneVision N series monitor).
(4)
Multi-pin connector: connects the N1 to the Modular Rack or Dock.
(5)
External DC power input connector: connects the N1 to the AC
adapter
(4)
(5)
(3)
(3)
(2)
(1)