15 - 2
BeneVision N1 Patient Monitor Operator’s Manual
15.2
CO
2
Safety Information
WARNING
•
Route all tubing away from the patient’s throat to avoid strangulation.
CAUTION
•
Remove the airway sample line from the patient’s airway while nebulized medications are being
delivered.
•
EtCO
2
values measured from the CO
2
module may differ from those of from the blood gas analysis.
•
Avoid mechanical shock to the sidestream CO
2
module
NOTE
•
The CO
2
module automatic suppress physiological alarms until breathing waves have been
detected. Make sure that a patient is properly connected when monitoring with the CO
2
module.
15.3
CO
2
Measurement Limitations
The following factors may influence the measurement accuracy:
■
Leaks or internal venting of sampled gas
■
Mechanical shock
■
Cyclic pressure up to 10 kPa (100 cmH
2
O)
■
Other sources of interference, if any
Measurement accuracy of the sidestream CO
2
module may be affected by the breath rate and inspiration/
expiration (I/E) ratio. Measurement accuracy of the microstream CO2 module may be affected by the breath
rate.For more information, refer to
(1) CO
2
menu hard key
(2) CO
2
Measure/standby hard key
(3) CO
2
watertrap seat
(4) Gas outlet
(5) Sample line connector
(6) CO
2
sensor connector
(7) CO
2
adapter connector
(7)