BeneVision N1 Patient Monitor Operator’s Manual
A - 15
PR from Nellcor SpO
2
Module
PR from Masimo SpO
2
Module
PR from NIBP Module
PR from IBP Module
Measurement range
20 to 300 bpm
Resolution
1 bpm
Response time
≤30 s (normal perfusion, no disturbance, PR value sudden change from 25 to
250 bpm)
Accuracy
20 to 250 bpm: ±3 bpm
251 to 300 bpm, not specified
Refreshing rate
≤1 s
Measurement range
25 to 240 bpm
Resolution
1 bpm
Response time
≤20 s (normal perfusion, no disturbance, PR value sudden change from 25 to
220 bpm)
Accuracy
±3 bpm (without motion)
±5 bpm (with motion)
Refreshing rate
≤2 s
Measurement range
30 to 300 bpm
Resolution
1 bpm
Accuracy
±3 bpm or ±3%, whichever is greater
Measurement range
25 to 350 bpm
Resolution
1 bpm
Accuracy
±1 bpm or ±1%, whichever is greater
Содержание BeneVision N1