14 - 4
BeneVision N1 Patient Monitor Operator’s Manual
1.
Disconnect the ICP adapter cable from the measurement module, or remove the module from the monitor.
2.
Connect the ICP adapter cable, measurement module, and the target monitor, or insert the measurement
module into the target monitor.
3.
Check that the zero reference value displayed on the monitor is consistent with that recorded on the ICP
transducer.
◆
Consistent: select
Accept
.
◆
Inconsistent: input the zero reference value recorded on the ICP transducer, and select
Accept.
CAUTION
•
If monitors of different brands are used to zero the Codman ICP transducer, the zero reference
values can be different. Use a Mindray monitor to Zero the Codman ICP transducer if you will take
ICP measurement using a Mindray monitor. Otherwise the ICP measurement can be inaccurate.
14.5
IBP Display
The IBP measurement is displayed on the monitor as a waveform and numeric pressures. For arterial pressure,
the IBP numeric area displays systolic pressure, diastolic pressure and mean pressure. For venous pressure, the
IBP numeric area displays only the mean pressure. The figure below shows the waveform and numerics for the
Art pressure.
14.6
Changing IBP Settings
14.6.1
Changing the IBP Alarm Settings
To change the IBP alarm settings, follow this procedure:
1.
Select the IBP numeric area or waveform area to enter the corresponding pressure menu.
2.
Select the
Alarm
tab.
3.
Enter the password if required. For more information, refer to
6.9.3 Selecting Password for User
.
4.
Set the alarm properties.
14.6.2
Changing the Pressure Label
The pressure label is a unique identifier for each type of pressure. Therefore, you should select a proper pressure
label for the source of the pressure you want to monitor.
To select the pressure label, follow this procedure:
1.
Select the IBP numeric area or waveform area to enter the corresponding pressure menu.
2.
Select the
Setup
tab.
(1) Pressure label
(2) Waveform
(3) Pressure Unit
(4) Systolic pressure
(5) Diastolic pressure
(6) Mean pressure
(7) PPV measurement
(7) (6)
(1) (2) (3) (4) (5)
Содержание BeneVision N1
Страница 1: ...BeneVision N1 Patient Monitor Operator s Manual...
Страница 273: ...F 8 DCI DCIP 4050 4051 LNCS YI 4053...