E - 4
BeneVision N1 Patient Monitor Operator’s Manual
E.8
Patient Auxiliary Current
Patient Auxiliary currents are measured between any selected Applied Part connector and the remaining
Applied Part connector s. All measurements may have a true RMS only response.
The following outlet conditions apply when performing the Patient Auxiliary Current test.
■
normal polarity (Normal Condition);
■
reverse polarity (Normal Condition),
■
normal polarity with open neutral (Single Fault Condition);
■
reverse polarity with open neutral (Single Fault Condition).
■
normal polarity with open earth (Single Fault Condition);
■
reverse polarity with open earth (Single Fault Condition).
LIMITS
For CF
applied parts,
◆
10 μA in Normal Condition
◆
50 μA in Single Fault Condition
For BF
applied parts,
◆
100 μA in Normal Condition
◆
500 μA in Single Fault Condition
NOTE
•
Make sure the safety analyzer is authorized comply with requirement of IEC60601-1.
•
Follow the instructions of the analyzer manufacturer.