7 - 10
BeneVision N1 Patient Monitor Operator’s Manual
When the alarm pause time expires, the alarm paused status is automatically deactivated. You can also cancel
the alarm paused status by pressing the
Alarm Pause
quick key.
7.7.2.1
Setting the Alarm Pause Time
The alarm pause time can be set to
1 min
,
2 min
,
3 min
, or
Permanent
. The default alarm pause time is
2 min
.
To set the alarm pause time, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required
password
→
select .
2.
Select the
Alarm
tab
→
Pause/Reset
tab.
3.
Set
Pause
Time
.
7.7.2.2
Prolonging the Alarm Pause Time
You can temporarily prolong the alarm pause time after the monitor enters the alarm paused status. This
function is enabled by default. To do so, follow this procedure:
1.
In the physiological alarm information area, select the alarm pause countdown.
2.
Select
Pause 5 min,
Pause 10 min
, or
Pause 15 min
.
To disable the function of prolonging the alarm pause time, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required
password
→
select .
2.
Select the
Alarm
tab
→
Pause/Reset
tab.
3.
Switch off
Pause 5 min
,
Pause 10 min
, or
Pause 15 min
.
NOTE
•
Prolonging alarm pause time does not affect the setting of alarm pause time.
•
You can not prolong the pause time if the Pause Time is set to Permanent.
7.7.2.3
Selecting the Priority of Paused Alarms
To select alarms of what priority can be paused, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required
password
→
select .
2.
Select the
Alarm
tab
→
Pause/Reset
tab.
3.
Set
Pause Priority
.
◆
All
: pressing the
Alarm Pause
quick key pauses all alarms.
◆
Med & Low
: pressing the
Alarm Pause
quick key pauses alarms of medium and low priority. The high
priority alarms will not be paused.
◆
Disabled
: the
Alarm Pause
quick key is disabled.
7.7.2.4
Switching Off All Alarms
If
Pause Time
is set to
Permanent
7.7.2.1 Setting the Alarm Pause Time
), pressing the
Alarm Pause
quick key
permanently switches off all alarms. The alarm off status has the following features:
■
Physiological alarms are switched off. The alarm lamp does not flash and alarm sound is not issued.
■
Alarm sound of technical alarms is switched off, but alarm lamp flashes and alarm messages are presented.
■
The message
Alarm Off
with red background is displayed in the physiological alarm information area.
■
The alarm off symbol is displayed in the system status information area.
To exit the alarm off status, press the
Alarm Pause
quick key again.