XZR400 Series User’s Manual
Michell Instruments
83
APPENDIX G
Appendix G
Commutable Scale (Auto-Ranging)
This option allows following the O
2
concentration:
•
Between 0 and 10 ppm on the
fi
rst 4-20 mA output
•
Between 0 and 100ppm, 0 and 1000ppm, 10 and 10000ppm, or
1 and 25% on the second 4-20 mA output
The alarm contacts 1 and 2 indicate the beginning and the end of the scale used on the
second 4-20 mA output.
Settings are indicated below:
4-20 mA
output
Low Scale
Up Scale
Alarm 1
Contact
Position
Alarm 2
Contact
Position
No 1
0
10ppm
No 2
0
100ppm
Open
Open
0
1000ppm
Closed
Open
10ppm
10000ppm
Open
Closed
1%
25%
Closed
Closed