MAINTENANCE
Observation
Cause
Corrective Action
Display
Indicates
T/C Fault
Open circuit
thermocouple
BEWARE OF ELECTRIC SHOCK / BURN HAZARD
• Remove the analyzer cover
• Check for faulty wiring of sensor - repair as
necessary
• Replace the MSRS / thermocouple sensor
assembly
• Recalibrate instrument
Broken thermocouple
BEWARE OF ELECTRIC SHOCK / BURN HAZARD
• Remove the analyzer cover
• Replace the MSRS / thermocouple sensor
assembly
• Recalibrate instrument
Faulty analyzer
Return the analyzer to Michell Instruments for further
investigation
Display
Indicates
Low Flow
Gas
fl
ow too low
BEWARE OF ELECTRIC SHOCK / BURN HAZARD
• Remove the analyzer cover
• Measure voltage across J11 pin 11 to J12 pin 10
• If 13 V < measured value < 15 V return the
analyzer to Michell Instruments for further
investigation
Leak in gas sample
path
BEWARE OF ELECTRIC SHOCK / BURN HAZARD
• Remove the analyzer cover
• Ensure all gas
fi
ttings are tight
• If the problem persists return the analyzer to
Michell Instruments for further investigation
Faulty electronic
fl
ow
meter
Return the analyzer to Michell Instruments for further
investigation
Faulty analyzer
Return the analyzer to Michell Instruments for further
investigation
Display
Indicates
High Flow
Gas
fl
ow too high
• Open the
fl
ow adjustment valve
• Close the bypass valve
• Adjust both valves to achieve a gas sample
fl
ow
to 2 l/hr ± 1 l/hr
Faulty electronic
fl
ow
meter
Return the analyzer to Michell Instruments for sensor
replacement
Faulty analyzer
Return the analyzer to Michell Instruments for further
investigation
Содержание XZR-400-BM
Страница 1: ...XZR400 Series Oxygen Analyzers User s Manual 97472 Issue 3 August 2018 ...
Страница 76: ...XZR400 Series User s Manual 66 97472 Issue 3 August 2018 APPENDIX A Appendix A Technical Specifications ...
Страница 79: ...XZR400 Series User s Manual Michell Instruments 69 APPENDIX B Appendix B Modbus RTU over RS485 ...
Страница 83: ...XZR400 Series User s Manual Michell Instruments 73 APPENDIX C Appendix C RS232 Serial Output ...
Страница 90: ...XZR400 Series User s Manual 80 97472 Issue 3 August 2018 APPENDIX F Appendix F Flow Fault Contact Optional ...
Страница 92: ...XZR400 Series User s Manual 82 97472 Issue 3 August 2018 APPENDIX G Appendix G Commutable Scale Auto Ranging ...
Страница 94: ...XZR400 Series User s Manual 84 97472 Issue 3 August 2018 APPENDIX H Appendix H Optional moisture sensor ...
Страница 101: ......
Страница 102: ...http www michell com ...