![Michell Instruments XZR-400-BM Скачать руководство пользователя страница 35](http://html.mh-extra.com/html/michell-instruments/xzr-400-bm/xzr-400-bm_user-manual_1784693035.webp)
XZR400 Series User’s Manual
Michell Instruments
25
INSTALLATION
2.10.3 Electrical Terminal Block
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15 16
Item Function
1
Connection to the MSRS cell; system reserved (TC+ – orange).
2
Connection to the MSRS cell; system reserved (common – white).
3
Connection to the MSRS cell; system reserved (reference – blue).
4-5
Oven connection, system reserved.
6
Optional (+) 4-20 mA output. Measurement proportional to the chosen O
2
scale.
A second 4-20 mA output is available as an option.
7
Optional 4-20 mA (0 V).
8-9
Optional Dry contact alarm for
fl
ow rate setting (250 V AC, 30 V DC or 2 A, 2 A
resistive load). Direction of action (positive, normal).
10-11 Optional General alarm dry contact terminal (250 V AC, 2 A or 30 V DC, 2 A on
resistive load). Direction of action (positive, normal).
12
Optional Threshold alarm No. 1 dry contact terminal (250 V AC , 2 A or 30 V DC,
2 A on resistive load). Function mode (positive, normal) and hysteresis can be
con
fi
gured by setting the parameters.
13
Optional Shared contact for alarms No.1 and No. 2.
14
Optional Threshold alarm No. 2 dry contact terminal (250 V AC, 5 A or 30 V DC,
5 A on resistive load). Function mode (positive, normal) and hysteresis can be
con
fi
gured by setting the parameters.
15-16 Not connected.
Figure 24
Electrical Terminal Block XZR400A3
Содержание XZR-400-BM
Страница 1: ...XZR400 Series Oxygen Analyzers User s Manual 97472 Issue 3 August 2018 ...
Страница 76: ...XZR400 Series User s Manual 66 97472 Issue 3 August 2018 APPENDIX A Appendix A Technical Specifications ...
Страница 79: ...XZR400 Series User s Manual Michell Instruments 69 APPENDIX B Appendix B Modbus RTU over RS485 ...
Страница 83: ...XZR400 Series User s Manual Michell Instruments 73 APPENDIX C Appendix C RS232 Serial Output ...
Страница 90: ...XZR400 Series User s Manual 80 97472 Issue 3 August 2018 APPENDIX F Appendix F Flow Fault Contact Optional ...
Страница 92: ...XZR400 Series User s Manual 82 97472 Issue 3 August 2018 APPENDIX G Appendix G Commutable Scale Auto Ranging ...
Страница 94: ...XZR400 Series User s Manual 84 97472 Issue 3 August 2018 APPENDIX H Appendix H Optional moisture sensor ...
Страница 101: ......
Страница 102: ...http www michell com ...