XZR400 Series User’s Manual
Michell Instruments
41
OPERATION
3.6
The Main (Expert) Menu
This menu accesses all of the con
fi
guration functions of the analyzer. Press the
corresponding area to display the required screen.
Flow
Analog output
Alarms
Auto adjustment
System
Comm
Pressure
Main menu
5
Figure 36
Main Menu Screen
Area
Function
Section
Analog output
Con
fi
gures the 4-20 mA analog output 1
3.6.1
Alarms
Disables the 3 alarms (during adjustment and/or during
normal operation) and sets the threshold and the function
mode for alarm 1 and 2
3.6.2
Auto adjustment
Con
fi
gures the automatic cyclical adjustment (adjust) of
the analyzer
Optional
3.6.3
System
Con
fi
gures the analyzer timestamp function, oven
temperature, menu display language, Expert access
code and RS output con
fi
guration (ModBus address with
RS485 or frame period with RS232).
3.6.4
Pressure
Con
fi
gures the process pressure correction
Optional for all models
3.6.5
Returns to the Measurement Display Screen (3)
COM 232
Sets the frequency of the frame transmission
Optional
Cannot coexist with the RS485
3.6.6
COM 485
Sets the ModBus address of the analyzer and displays
the message frames received by the analyzer via the
RS485 interface
Cannot coexist with the RS232
3.6.6
Flow
Con
fi
gures the gas sample
fl
ow correction based on its
density
3.6.7
Содержание XZR-400-BM
Страница 1: ...XZR400 Series Oxygen Analyzers User s Manual 97472 Issue 3 August 2018 ...
Страница 76: ...XZR400 Series User s Manual 66 97472 Issue 3 August 2018 APPENDIX A Appendix A Technical Specifications ...
Страница 79: ...XZR400 Series User s Manual Michell Instruments 69 APPENDIX B Appendix B Modbus RTU over RS485 ...
Страница 83: ...XZR400 Series User s Manual Michell Instruments 73 APPENDIX C Appendix C RS232 Serial Output ...
Страница 90: ...XZR400 Series User s Manual 80 97472 Issue 3 August 2018 APPENDIX F Appendix F Flow Fault Contact Optional ...
Страница 92: ...XZR400 Series User s Manual 82 97472 Issue 3 August 2018 APPENDIX G Appendix G Commutable Scale Auto Ranging ...
Страница 94: ...XZR400 Series User s Manual 84 97472 Issue 3 August 2018 APPENDIX H Appendix H Optional moisture sensor ...
Страница 101: ......
Страница 102: ...http www michell com ...