XZR400 Series User’s Manual
34
97472 Issue 3, August 2018
OPERATION
3.3 Warm-Up
Period
Upon power up, the screen appears as shown below. The analyzer performs a series of
internal checks for about 5 seconds.
NOTE: Touch the French or English area to display the menu in the required
language.
French
1
Software version V0.1.7.ver
English
Figure 29
Start-up Screen
Wait about 15 minutes until the oven temperature reaches 634°C, indicated by the oven
temperature line.
The low temperature alarm is displayed digitally throughout the rise in oven temperature;
the general alarm contact is activated.
NOTE: A fl ow greater than 3.5 l/h prevents the correct rise in oven temperature.
Oven
Flow
Low temperature alarm
48.8ºC
7.5 l/h
Figure 30
Oven Temperature Screen
Содержание XZR-400-BM
Страница 1: ...XZR400 Series Oxygen Analyzers User s Manual 97472 Issue 3 August 2018 ...
Страница 76: ...XZR400 Series User s Manual 66 97472 Issue 3 August 2018 APPENDIX A Appendix A Technical Specifications ...
Страница 79: ...XZR400 Series User s Manual Michell Instruments 69 APPENDIX B Appendix B Modbus RTU over RS485 ...
Страница 83: ...XZR400 Series User s Manual Michell Instruments 73 APPENDIX C Appendix C RS232 Serial Output ...
Страница 90: ...XZR400 Series User s Manual 80 97472 Issue 3 August 2018 APPENDIX F Appendix F Flow Fault Contact Optional ...
Страница 92: ...XZR400 Series User s Manual 82 97472 Issue 3 August 2018 APPENDIX G Appendix G Commutable Scale Auto Ranging ...
Страница 94: ...XZR400 Series User s Manual 84 97472 Issue 3 August 2018 APPENDIX H Appendix H Optional moisture sensor ...
Страница 101: ......
Страница 102: ...http www michell com ...