XZR400 Series User’s Manual
20
97472 Issue 3, August 2018
INSTALLATION
2.9.2 Electrical Terminal Block
The terminal block is located by removing the front panel cover.
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15 16
Item Function
1
Connection to the MSRS cell; system reserved (TC+ – orange).
2
Connection to the MSRS cell; system reserved (common – white).
3
Connection to the MSRS cell; system reserved (reference – blue).
4
Oven connection, system reserved.
5
Oven connection, system reserved (see paragraph 15.2).
6
(+) 4-20 mA output measurement proportional to the chosen O
2
scale. A second
4-20 mA output is available as an option.
7
4-20 mA (0 V).
8-9
Optional. Dry contact alarm for
fl
ow rate setting (250 V AC, 30 V DC or 2 A, 2 A
resistive load). Direction of action.
10-11 Threshold alarm No. 1 dry contact terminal (250 V AC , 2 A or 30 V DC, 2 A on
resistive load). Function mode (positive, normal) and hysteresis can be con
fi
gured
by setting the parameters.
12
Threshold alarm No. 1 dry contact terminal (250 V AC , 2 A or 30 V DC, 2 A on
resistive load). Function mode (positive, normal) and hysteresis can be con
fi
gured
by setting the parameters.
13
Shared contact for alarms No.1 and No. 2.
14
Threshold alarm No. 2 dry contact terminal (250 VAC , 5A or 30 VDC, 5A on
resistive load). Function mode (positive, normal) and hysteresis can be con
fi
gured
by setting the parameters.
15
Mains supply phase.
16
Mains supply neutral.
The equipment is protected by a T2 A - 250 V AC timed fuse (5 x 20 mm) located near
the terminal block
Figure 20
Electrical Terminal Block XZR400A2
2.9.3 D-Sub DE9 plug
This optional D-Sub DE9 type male plug (9 pin) is used to connect an RS485 or RS232
port.
Содержание XZR-400-BM
Страница 1: ...XZR400 Series Oxygen Analyzers User s Manual 97472 Issue 3 August 2018 ...
Страница 76: ...XZR400 Series User s Manual 66 97472 Issue 3 August 2018 APPENDIX A Appendix A Technical Specifications ...
Страница 79: ...XZR400 Series User s Manual Michell Instruments 69 APPENDIX B Appendix B Modbus RTU over RS485 ...
Страница 83: ...XZR400 Series User s Manual Michell Instruments 73 APPENDIX C Appendix C RS232 Serial Output ...
Страница 90: ...XZR400 Series User s Manual 80 97472 Issue 3 August 2018 APPENDIX F Appendix F Flow Fault Contact Optional ...
Страница 92: ...XZR400 Series User s Manual 82 97472 Issue 3 August 2018 APPENDIX G Appendix G Commutable Scale Auto Ranging ...
Страница 94: ...XZR400 Series User s Manual 84 97472 Issue 3 August 2018 APPENDIX H Appendix H Optional moisture sensor ...
Страница 101: ......
Страница 102: ...http www michell com ...