XZR400 Series User’s Manual
62
97472 Issue 3, August 2018
MAINTENANCE
5 MAINTENANCE
5.1
Troubleshooting Guide / Failure Analysis
Observation
Cause
Corrective Action
Mains power
plugged in and
switched on
but no display
present
Fuse blown
• Replace mains fuse:
• Fuse Box: T2A/250V
• Mains socket: T6.3A/250
Analyzer not powered
BEWARE OF ELECTRIC SHOCK / BURN HAZARD
• Remove the analyzer cover
• Measure the main inlet voltage
• If the measured voltage is 90 to 264 V AC verify
the mains power to the analyzer
Display disconnected
BEWARE OF ELECTRIC SHOCK / BURN HAZARD
• Remove the analyzer cover
• Ensure the plugs of the 20-way ribbon cable are
connected to both the display PCB and the mother
board and that the wiring is intact
• Ensure the plugs of the 2-way cable are connected
to both the display PCB and the mother board and
that the wiring is intact
• Ensure the plugs of the 4-way cable are connected
to both the display PCB and the mother board and
that the wiring is intact
Faulty analyzer
Return the analyzer to Michell Instruments for further
investigation
Содержание XZR-400-BM
Страница 1: ...XZR400 Series Oxygen Analyzers User s Manual 97472 Issue 3 August 2018 ...
Страница 76: ...XZR400 Series User s Manual 66 97472 Issue 3 August 2018 APPENDIX A Appendix A Technical Specifications ...
Страница 79: ...XZR400 Series User s Manual Michell Instruments 69 APPENDIX B Appendix B Modbus RTU over RS485 ...
Страница 83: ...XZR400 Series User s Manual Michell Instruments 73 APPENDIX C Appendix C RS232 Serial Output ...
Страница 90: ...XZR400 Series User s Manual 80 97472 Issue 3 August 2018 APPENDIX F Appendix F Flow Fault Contact Optional ...
Страница 92: ...XZR400 Series User s Manual 82 97472 Issue 3 August 2018 APPENDIX G Appendix G Commutable Scale Auto Ranging ...
Страница 94: ...XZR400 Series User s Manual 84 97472 Issue 3 August 2018 APPENDIX H Appendix H Optional moisture sensor ...
Страница 101: ......
Страница 102: ...http www michell com ...