XZR400 Series User’s Manual
Michell Instruments
79
APPENDIX E
Appendix E
Process Pressure Correction - (Optional)
When the process gas pressure is out of the atmospheric pressure range, the process
pressure correction is required to guarantee the optimal ef
fi
ciency of the analyzer.
The input signal is a 4-20 mA analog signal from an external pressure transmitter
installed by the user at the outlet side of the process.
The following screen is used to con
fi
gure the scale range:
Pressure correction
5.6
Start of scale
End of scale
0.0 bar abs
2.0 bar abs
Figure 54
Pressure Correction Screen
An input signal of 4 mA corresponds to the low scale set value.
A input signal of 20 mA corresponds to the high scale set value.
E.1
Process Pressure Correction Input Connections
The 4-20 mA pressure transmitter signals should be connected to the DB15 socket on
the rear panel of the analyzer.
1 2
9 10 11 12 13 14 15
3 4 5 6 7 8
Contact No
Function
5
(+) 4-20 mA / analog input of process pressure
6
(+) 4-20 mA / analog input of process pressure
Содержание XZR-400-BM
Страница 1: ...XZR400 Series Oxygen Analyzers User s Manual 97472 Issue 3 August 2018 ...
Страница 76: ...XZR400 Series User s Manual 66 97472 Issue 3 August 2018 APPENDIX A Appendix A Technical Specifications ...
Страница 79: ...XZR400 Series User s Manual Michell Instruments 69 APPENDIX B Appendix B Modbus RTU over RS485 ...
Страница 83: ...XZR400 Series User s Manual Michell Instruments 73 APPENDIX C Appendix C RS232 Serial Output ...
Страница 90: ...XZR400 Series User s Manual 80 97472 Issue 3 August 2018 APPENDIX F Appendix F Flow Fault Contact Optional ...
Страница 92: ...XZR400 Series User s Manual 82 97472 Issue 3 August 2018 APPENDIX G Appendix G Commutable Scale Auto Ranging ...
Страница 94: ...XZR400 Series User s Manual 84 97472 Issue 3 August 2018 APPENDIX H Appendix H Optional moisture sensor ...
Страница 101: ......
Страница 102: ...http www michell com ...