XZR400 Series User’s Manual
Michell Instruments
45
OPERATION
3.6.3 Auto Adjustment (Optional)
Purpose
This function is used for the analyzer’s auto adjust at a set time and at regular intervals
set in hours or days. Solenoid valves connected to the D-Sub DA15 on the XZR400A2,
and 8 pin connector on the XZR400A1 control any adjustment gas and control gas
inputs. Additional contacts are used to launch the sequence.
Hardware confi guration
Peripheral equipment must be installed as indicated.
Parameter setting for auto adjust
The Auto Adjust Screen sets the array of parameters used for the analyzer's auto adjust.
•
Touch the area to select it and change it.
•
Enter the new numeric value on the virtual keypad.
• Click
OK
to con
fi
rm or
Cancel
to discard the changes.
Main
Level
Auto adjustment
5.4.1
External trigger
T0 adj
Adjustment period
Unit
No
12:00
1 d
d
Process
Enable
Start
Process
Figure 40
Auto Adjustment Screen
The three tabs, Main, Level and Process are described next.
Содержание XZR-400-BM
Страница 1: ...XZR400 Series Oxygen Analyzers User s Manual 97472 Issue 3 August 2018 ...
Страница 76: ...XZR400 Series User s Manual 66 97472 Issue 3 August 2018 APPENDIX A Appendix A Technical Specifications ...
Страница 79: ...XZR400 Series User s Manual Michell Instruments 69 APPENDIX B Appendix B Modbus RTU over RS485 ...
Страница 83: ...XZR400 Series User s Manual Michell Instruments 73 APPENDIX C Appendix C RS232 Serial Output ...
Страница 90: ...XZR400 Series User s Manual 80 97472 Issue 3 August 2018 APPENDIX F Appendix F Flow Fault Contact Optional ...
Страница 92: ...XZR400 Series User s Manual 82 97472 Issue 3 August 2018 APPENDIX G Appendix G Commutable Scale Auto Ranging ...
Страница 94: ...XZR400 Series User s Manual 84 97472 Issue 3 August 2018 APPENDIX H Appendix H Optional moisture sensor ...
Страница 101: ......
Страница 102: ...http www michell com ...