XZR400 Series User’s Manual
16
97472 Issue 3, August 2018
INSTALLATION
XZR400A1 WITH INTERNAL PUMP
:
• 3 Swagelok 6mm stainless steel bulkhead unions (1 gas inlet and 2
gas outlets) - on the rear panel
• 1 sampling pump (3 l/min)
• 2
electrovalves
XZR400A1
Sensor flow adjustment valve
O
2
sensor
By-pass flow adjustment valve
Barometric pressure sensor
Ɨ
” bulkhead union
Ɨ
” bulkhead union
Ɨ
” bulkhead union
Sampling gas inlet
Electronic flowmeter 0-12 l/h
Sampling gas outlet
Sampling gas outlet
P
N.O. Solenoid valve
N.C. Solenoid valve
Sampling pump
Figure 17
Sample Path with Pump Fitted XZR400A1