XZR400 Series User’s Manual
58
97472 Issue 3, August 2018
CALIBRATION
Screen page 3.2.3 shows the O
2
concentration in the adjustment gas and the time left
before the purge sequence.
Cancel
Op forced
after 1min
Adjustment progressing
3.2.3
8.02 %
19 min 58 s
12.4 %
2.5 L/h
Standard gas :
remaining time :
Concentration O2
Flow
Adjustment
Figure 48
Screen Page 3.2.3
During the calibration sequence it is possible to:
•
Force calibration after 1 minute if the reading is stable. Press the
Op
forced after 1min
key and screen page 3.2.4 will show (after 1 minute).
•
Cancel calibration at any time. Press the
Cancel
key and go to screen
page 3.2.8.
Cancel Calibration
Cancel
bleeding
Bleeding running
3.2.8
29 min 58 s
8.02 %
remaining time :
Concentration :
Adjustment
Figure 49
Screen Page 3.2.8
Pressing the
Cancel bleeding
key will return to the Main Screen.
Содержание XZR-400-BM
Страница 1: ...XZR400 Series Oxygen Analyzers User s Manual 97472 Issue 3 August 2018 ...
Страница 76: ...XZR400 Series User s Manual 66 97472 Issue 3 August 2018 APPENDIX A Appendix A Technical Specifications ...
Страница 79: ...XZR400 Series User s Manual Michell Instruments 69 APPENDIX B Appendix B Modbus RTU over RS485 ...
Страница 83: ...XZR400 Series User s Manual Michell Instruments 73 APPENDIX C Appendix C RS232 Serial Output ...
Страница 90: ...XZR400 Series User s Manual 80 97472 Issue 3 August 2018 APPENDIX F Appendix F Flow Fault Contact Optional ...
Страница 92: ...XZR400 Series User s Manual 82 97472 Issue 3 August 2018 APPENDIX G Appendix G Commutable Scale Auto Ranging ...
Страница 94: ...XZR400 Series User s Manual 84 97472 Issue 3 August 2018 APPENDIX H Appendix H Optional moisture sensor ...
Страница 101: ......
Страница 102: ...http www michell com ...