XZR400 Series User’s Manual
30
97472 Issue 3, August 2018
INSTALLATION
2.12.2 Alarm Output Connections
Two alarm relays are provided and are connected to the instrument via the terminal
block on the back panel of the analyzer.
The signal outputs will be connected to external systems that can
potentially infl uence the operation of the process.
The alarm level signals could also be at mains potential so it is
essential that, before connecting these signal lines, checks are
made to ensure that these inputs are not live and that it is safe to
handle them.
Alarms 1 and 2 are concentration alarms. The direction of the activation N/O or N/C, as
well as the hysteresis, can be con
fi
gured.
The output contacts are normally open and potential free. The relay switching capability
is 10 W max (up to 100 V or up to 0.5 A).
The method of connection is as follows:
Always use screened cable to connect the output to the external
device.
Alarm 1
1. Strip back the two wires for the alarm 1 output leads, exposing
approximately 6mm (0.25") wire and clamp them into the two screw ports
labelled as
Alarm 1
.
Do not overtighten the screw.
2.
Connect the screen to the ground port.
Alarm 2
1. Strip back the two wires for the alarm 2 output leads, exposing
approximately 6mm (0.25") wire and clamp them into the two screw ports
labelled as
Alarm 2
.
Do not overtighten the screw.
2.
Connect the screen to the ground port.
Содержание XZR-400-BM
Страница 1: ...XZR400 Series Oxygen Analyzers User s Manual 97472 Issue 3 August 2018 ...
Страница 76: ...XZR400 Series User s Manual 66 97472 Issue 3 August 2018 APPENDIX A Appendix A Technical Specifications ...
Страница 79: ...XZR400 Series User s Manual Michell Instruments 69 APPENDIX B Appendix B Modbus RTU over RS485 ...
Страница 83: ...XZR400 Series User s Manual Michell Instruments 73 APPENDIX C Appendix C RS232 Serial Output ...
Страница 90: ...XZR400 Series User s Manual 80 97472 Issue 3 August 2018 APPENDIX F Appendix F Flow Fault Contact Optional ...
Страница 92: ...XZR400 Series User s Manual 82 97472 Issue 3 August 2018 APPENDIX G Appendix G Commutable Scale Auto Ranging ...
Страница 94: ...XZR400 Series User s Manual 84 97472 Issue 3 August 2018 APPENDIX H Appendix H Optional moisture sensor ...
Страница 101: ......
Страница 102: ...http www michell com ...