-65-
Errors and Problems
3-013 Warning: „mesuNET: No connection to the server“
The log data (mesuNET) buffered on the unit concerned exceeds a size of 3 KB (approx. 40 messages).
Check the Link0 -LED. Check the connection.
3-014 Error: „mesuNET: No connection to the server –
loss of data“
The log data (mesuNET) buffered on the unit concerned exceeds a size of 100 KB (approx. 1300 messages).
In this case no further data will be buffered.
Check the Link0 -LED. Check the connection.
3-015 Error: „Detector test failed: timeout“
No
metal signal was detected over the duration of the detector test.
see chapter
Adjustment
Operating level QM
Detector test
3-016 Error: „Detector test failed: Tolerance“
This message appears if the result of the comparison of the determined test body and the learned test body
(with assigned tolerances) is negative.
3-017 Warning: „Password expired“
Please assign a new password.
see chapter
Adjustment
Operating level 1
Change password
3-018 Notice: „Protocol size near maximum“
The size of the log database exceeds 19500 entries. Please backup and delete logs. The log database is
limited to 20000 entries.
see chapter
Settings
Operating level QM
Protocol
3-019 Warning: „Protocol size at maximum“
The size of the log database has reached 20000. 1000 old entries will be deleted.
The log database is limited to 20000 entries.
see chapter
Settings
Operating level QM
Protocol