-50-
Adjustment
Operating level QM
User Manager
New
The
Name/Access code
tab is accessed via ‘New’. Enter your user name and a 4-digit access code via the
keyboard. On the
Rights
tab, select the
levels
accessible to the user and whether the respective user may
abort a detector test or whether he may confirm a failed detector test (
Reset detector test
). The user is
created with ‘Save’.
see chapter
Adjustment
Operating level QM
Detector test
Code entry required
Edit
Select the user and click on ‘Edit’.
Delete
Select the user and click on ‘Delete’.
Password expires
If ‘yes’, each user will be requested to change his password after the expiry of the indicated number of days
(standard: 42 days).
see chapter
Adjustment
Operating level QM
Password expires in
Password expires in (days) (optional)
This menu item is displayed only if ‘password expires’ has been confirmed with ‘yes’.
The password can be valid for a period of 1 to 365 days.
see chapter
Adjustment
Operating level QM
Password expires
Protocol
The evaluation electronics continuously record all occurring metal events and errors together with the date
and time. You have the option to export the protocol with the help of the USB connection.
You can select via a button whether the database is to be exported as a text file (file extension .txt) or as
‘comma separated values’ (file extension .csv).
Example of a csv export:
ID
Type
Timestamp
Number
Unit
Text
Batchnumber
Productnumber
Productname
User
1 notice 11.10.2012
14:46
901
1
Device on: 2012-10-11
14:46:58 2012-01 1
Product
1
QM
2 notice 11.10.2012 14:50
911
1 Logged-in user: QM
2012-01
1 Product 1
QM
3 notice 11.10.2012
16:24
913
1
Detection threshold changed
from 80 to 75
2012-01
1 Product 1
QM
4 notice 11.10.2012 17:02
900
1 Metal 865 59.8°
2012-01
1 Product 1
QM
5 notice 11.10.2012 18:36
912
1 Load product -> product 2
------------
2 Product 2
QM
Содержание METRON 07 C