+49 8142 448 460
www.medin-medical.com
medin Medical Innovations GmbH
Adam-Geisler-Str. 1
D-82140 Olching
OP_medinCNO EN
Rev 12 Stand 06.03.2019
56
von
68
perform this together with reactivation of the oscillation. (Exception: Only these alarms were active:
Disconnection, CPAP pressure high, Feed pressure high
or
Press Alarm Reset to restart flow
)
-
If the emergency programme is triggered by a
CPAP pressure high
or
Feed pressure high
alarm, the
medinCNO
®
stops its flow. If the triggering high pressure decreases immediately (<200 msec), the
medinCNO
®
restarts its flow after approximately 4 seconds. If it takes longer than this amount of
time for the pressure to decrease, the medinCNO
®
restarts its flow gradually, whereby it takes
approximately 10 seconds for the medinCNO
®
to once again emit its set flow. If the alarm was
triggered by a blocked flow supply tube or a combination of several alarms of the
CPAP pressure
high
or
feed pressure high
type, the flow is stopped for the time being and in addition to the
feed
pressure high/CPAP pressure high
alarm, a text message appears, informing the user that the flow
will be restarted by pressing the
Alarm Reset
button.
-
If the emergency programme is caused by the breakdown of one of the supply gases, the
medinCNO
®
continues its basic CPAP operation with the remaining gas so that an oxygen
concentration of 21% or 100% develops.
-
If there is a serious error in the device, the flow is deactivated and the device must be restarted.
8.4
Indicator lights
The medinCNO
®
contains two LEDs with the following meanings:
Alarm LED:
Colour
Meaning
Red
A highest-priority alarm is present – Immediate action by the operator is necessary.
Yellow
A medium- or low-priority alarm is present – Prompt action by the operator is
required.
White
Information signal: two short blinks indicate that a push was triggered. This signal can
be deactivated in many situations by the operator (see chapter 5.12) as needed.
LED in the dial:
This LED is activated if the operator can change or confirm settings by turning or pressing the dial. If this LED is
deactivated, the dial does not have any function at the present moment.