Warning:
The standby mode can only be used when no patient is attached to the medinCNO
®
.
Please note:
In standby mode, no gas flow can be emitted from the device. When combining with an active respiratory gas
humidifier, the following points must be borne in mind:
-
If no gas flow is issued and if the respiratory gas humidifier is continuing to heat, there can be so-
called "hot shots". This describes the brief delivery of overly hot respiratory gas to the patient
which collected in the respiratory gas humidifier chamber during time spent in standby mode.
Recommendation:
In the case of a brief period in standby mode (manipulation such as bonnets and/or mask/prong
changes, suctioning, etc.) and activated humidifier, the medinCNO
®
should be restarted before
the patient is connected to the device. This will avoid the emission of hot shots to the patient.
In the case of a longer period in standby mode (pause in CPAP therapy, skin-to-skin care, etc.),
the respiratory gas humidifier should be switched off.
-
In the case of a longer period in standby mode and deactivated respiratory gas humidifier, the
tubing system should be hung as vertically as possible so that any condensate can flow back into
the respiratory gas humidifier. The Medijet
®
generator should be neatly covered. In addition, the
7-day rule regarding replacement of the tubing system applies.