The internal 3-volt battery must be replaced as required or every three years at the latest. The
replacement battery must be a CR123A powerone from VARTA Microbattery GmbH (REF 39-102).
-
The internal rechargeable battery must be replaced as required or every three years at the latest.
The replacement battery must be a rechargeable battery of the same type (can be ordered from
Medin, REF 39-101).
-
If the medinCNO
®
contains a water trap, the filter contained within it (REF 39-819) has to be checked
regularly and changed at the latest after one year.
-
If the medinCNO
®
contains a prefilter in the ventilation of the housing, the filter contained within it
(REF 39-317) has to be checked regularly, cleaned if necessary and changed at the latest after one
year.
If an error is detected during this function test and maintenance, the device must not be used or connected to
a patient under any circumstances.
Warning:
The two internal batteries (3V and rechargeable) may only be replaced by the battery types listed above and
only by trained, professional service personnel and in accordance with the instructions in the service manual.
After each maintenance or repair of the medinCNO
®
, a complete function test must be performed and passed
before this medinCNO
®
can be used on a patient once again.
9.3
Repairs
Warning:
Repairs may only be made to the medinCNO
®
by specially trained, professional service personnel in accordance
with the instructions and warnings in the service manual. After each maintenance or repair of the medinCNO
®
,
a complete function test must be performed and passed before this medinCNO
®
can be used on a patient once
again.
9.4
Device modifications
Warning:
The medinCNO
®
may not be modified without permission from the manufacturer and subsequent appropriate
examinations and testing to ensure continued safe use.
9.5
Power supply unit (cleaning, maintenance, repairs and modifications)
Cleaning:
Prior to cleaning, the power supply unit of the medinCNO
®
must be disconnected from the mains.
Do not clean with chemical cleaning agents.
Maintenance/repairs/modifications:
The power supply unit FW7405M/24 of the medinCNO
®
is maintenance-free. It should not be opened.